Flexural Testing Acrylic Resin Cements Used in Orthopaedics (ISO 16402)

During the implantation process, bone cements are used to fill the space between the implant and the bone, and their long-term mechanical stability is essential for success of the implant. By undertaking static and cyclic loading, the mechanical performance of these bone cements can be established.

ISO 16402 ‘Implants for surgery – Acrylic resin cement - Flexural fatigue testing of acrylic resin cements used in orthopaedics’ provides a guideline for flexural testing of these cements in a laboratory situation. It consists of a method for both quasi-static bend testing and flexural fatigue.

A miniature 4-point bend fixture is used within into a fluid bath filled with Ringer’s solution and maintained at 37° C. After a quasi-static bend test is performed, fatigue testing is carried out to generate a Stress level/Number of cycles (S/N) curve.

We recommend the use of either an ElectroPuls™ All-Electric Dynamic Test Instrument or 8870 servohydraulic fatigue test system for undertaking both the quasi-static and the cyclic loading of these bone cements.

We recommend that you review the standard to fully understand its requirements.

Instron has a variety of solutions for testing of orthopaedic implants and devices or biomaterials. Find out more.


Über diese Lösung

Art der Prüfung: ISO 16402
Art der Probe: rechteckiger Querschnitt
Materialien: Biomedizin | Biomaterialien
Nach Material/Branche Biegeprüfung | Langzeit-Schwingfestigkeitsprüfung
Geschäftsfeld: Wissenschaft | Biomedizin, Medizinprodukte, Healthcare
4 Point Submersible Bend Fixture
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
Sind Sie an dieser Lösung interessiert?
Kontakt

Literatur zum Thema

ElectroPuls™ elektrodynamische Prüfsysteme   ElectroPuls™ elektrodynamische Prüfsysteme
ORTHOPÄDIE & OSTEOSYNTHESEVORRICHTUNGEN   ORTHOPÄDIE & OSTEOSYNTHESEVORRICHTUNGEN
Serie 5900 | Zuverlässige Prüfergebnisse   Serie 5900 | Zuverlässige Prüfergebnisse
Solutions for Biomedical Testing   Solutions for Biomedical Testing

Unsere Laboratorien

Die Instron Anwendungslaboratorien beschäftigen sich nicht nur mit Forschung und Entwicklung, sondern unterstützen auch durch Anwendungsbeispiele den Kunden bei der Auswahl des optimalen Prüfsystems.
Wenden Sie sich an unsere Anwendungsexperten und finden Sie